1. Product Application:
This specification is dedicated to medical nebulization therapy equipment. Its working principle is based on the PZT piezoelectric effect. The diaphragm is precisely drilled with micro-holes of 2 to 3 microns in diameter using high-precision laser drilling technology. Under the excitation of the electric field generated by the driving PCBA, the stainless steel diaphragm triggers high-frequency resonance, causing it to vibrate. This vibration atomizes the liquid medicine into aerosol particles ranging from 1 to 3.1 microns. These particles directly act on the lower respiratory tract or alveolar surface for deposition, achieving the purpose of treating asthma and respiratory diseases.
II. Product Model: / Type: φ16-110KSDB300
III. Performance Specifications
1 | 谐振频率Fr / Resonant Frequency | KHz | 110±5 |
2 | 谐振阻抗 / Resonant Impedance | Ω | < 200 |
3 | 静电容量Ct/ Static Capacitance | PF | 3000± 10% |
4 | 雾化片最大电压 / Input Voltage | Vpp | 80 |
5 | 雾化片功率 / Power | W | 1.5 ±0.5 |
6 | 使用寿命 | H | 1000 |
7 | 中位粒径 /MMAD(上下限制) | um | < 3.5 |
8 | 雾化量(雾化速率根据功率大小而变化) | ml/min | > 0.20 |
9 | 使用水源 / Using Water Quality | 盐水或者饮用水Salt water and drinking water |

IV. Specific parameters: Product material: SUS316L + PZT + nylon wire; pore size: 3±0.50um
Number of holes: 2000-3500 holes (the number of holes can be quantified, optimized, increased, decreased, and adjusted according to product technical requirements) Hole ratio: 95%
Wire: The length of the wire is customized according to customer requirements
V. Testing methods
5.1 Test Environment: 23±5℃, 40 ~ 70%RH
5.2 Test Instrument:
4.2.1 Tsinghua University Impedance Analyzer PV70A
4.2.2 Median diameter particle analyzer (for detecting median mass air dynamic diameter (MMAD) of atomized particles): WN-311XP
4.2.3 Atomization quantity test bench: Customized atomization quantity testing equipment, including electronic weighing.
VI. Precautions
Thank you for choosing our atomizing chip. To safeguard your company's interests, please ensure you carefully read the product precautions and the technical acknowledgment letter for the atomizing chip signed with your company, and keep them properly as proof for future services.
1. This product is strictly prohibited from being used in dry burning mode, and it is also prohibited from being used with an overpower of more than 1.5W. However, using it with less than 1W is also not suitable for atomization, as it may cause water accumulation during atomization.
2. In the structural design, the atomizing plate should be installed firmly and flatly. It is recommended to use food-grade silicone rubber rings with a hardness of 30~35 degrees for waterproofing.
3. Pressing is strictly prohibited within the area with a diameter of 14mm, otherwise it will affect the output of atomization volume and the sensory effect during atomization.
4. The connecting wire of the atomizing plate cannot be re-welded, otherwise it may cause damage to the atomizing plate;
5. The atomizing plate contains ceramic material, so avoid applying heavy pressure or subjecting it to strong vibrations or impacts.
6. It is strictly prohibited to use sharp or hard objects to scratch or prick the atomizing area of the steel sheet. Otherwise, the effect of the atomizing sheet may be altered or it may be directly damaged.
OUR ADVANTAGE
High Customizability
Ultra-fine Atomized Particles
Excellent Stability
Low Power Consumption Design
Strong Production Capacity